2012 Fiscal Year Final Research Report
Monitoring of the blood anesthetics gas levels using interference enhanced reflection method
Project/Area Number |
22659280
|
Research Category |
Grant-in-Aid for Challenging Exploratory Research
|
Allocation Type | Single-year Grants |
Research Field |
Anesthesiology/Resuscitation studies
|
Research Institution | Shiga University of Medical Science |
Principal Investigator |
|
Co-Investigator(Kenkyū-buntansha) |
KITAGAWA Hirotoshi 滋賀医科大学, 医学部, 講師 (50252391)
|
Project Period (FY) |
2010 – 2012
|
Keywords | 干渉増幅反射式センサー / 揮発性麻酔薬 / モニタリング / プロポフォール ベッドサイド |
Research Abstract |
Although the monitoring of blood anesthetic gas levels is an important for the safety of general anesthesia, no bedside monitoring method was found. We tested to monitor its levels using the interference enhanced reflection method with head space technique. It could get a quick response to the blood levels of isoflurane (0-5%), sevoflurane(0-5%) and desflurane(0-18%). Additionally, it allowed to monitor by only 1 ml blood sample volume by improving the head space technique.
|