2012 Fiscal Year Final Research Report
Standardization of patient dosimetry and investigation of the efficacy of radioprotective agent during neuroendovascular procedure
Project/Area Number |
22791231
|
Research Category |
Grant-in-Aid for Young Scientists (B)
|
Allocation Type | Single-year Grants |
Research Field |
Radiation science
|
Research Institution | 公益財団法人冲中記念成人病研究所 (2012) Okinaka Memorial Institute for Medical Research (2010-2011) |
Principal Investigator |
HAYAKAWA Mikito 公益財団法人冲中記念成人病研究所, 研究員 (80450229)
|
Project Period (FY) |
2010 – 2012
|
Keywords | 蛍光ガラス線量計 / 皮膚透過線量 / インターベンショナル・ラジオロジー / 医療被ばく / 脳血管内治療 |
Research Abstract |
We established a direct dosimetry method of patients’ entrance skin doses during neuroendovascular procedure using multiple radiophotoluminescence glass dosimeters. We applied this dosimetry system in clinical neuroendovascular settings, and clarified the frequency of maximum entrance skin dose (ESD) over 3 Gy, the prevalence of temporary epilation as a radiation-induced skin injury, and the difference of radiation-dose distributions between single-plane and bi-plane angiography equipments. In addition, we clarified that total fluoroscopic time and dose-area products were clearly related to maximum ESD and we could estimate the ESD in real-time during neuroendovascular procedure. These results could be the fundamental data to reduce radiation exposure during neuroendovascular procedure and to apply the radioprotective agent in a clinical setting in the near future.
|
-
[Journal Article] Precise mapping system of entrance skin dose during endovascular embolization for cerebral aneurysm2011
Author(s)
Moritake T, Hayakawa M, Matsumaru Y, Takigawa T, Koguchi Y, Miyamoto Y, Mizuno Y, Chida K, Akahane K, Tsuboi K, Sakae T, Sakurai H.
-
Journal Title
Radiation Measurements
Volume: 46
Pages: 2103-2106
-
-
-
[Presentation] Difference of skin dose distribution between bi-plane and single-plane systems during neurointerventional procedures2012
Author(s)
Hayakawa M, Moritake T, Matsumaru Y, Kawauchi S, Koguchi Y, Miyamoto Y, Akahane K, Sakae T, Tsuboi K, Sakurai H
Organizer
2012 World Congress on Medical Physics and Biomedical Engineering
Place of Presentation
Beijing, China
Year and Date
20120526-31
-
-
-
[Presentation] Precise mapping of entrance skin dose during neurointerventional procedure: what is the most effective method to avoid radiation-induced skin injury?2010
Author(s)
Moritake T, Hayakawa M, Matsumaru Y, Koguchi Y, Miyamoto Y, Mizuno Y, Chida K, Akahane K, Tsuboi K, Sakae T, Sakurai H.
Organizer
The 56th Annual Meeting of Radiation Research Society in USA
Place of Presentation
Maui, Hawaii, USA
Year and Date
20100926-29
-
-
[Presentation] Precise mapping of entrance skin dose during endovascular embolization for cerebral aneurysm to avoid radiation-induced skin injury2010
Author(s)
Moritake T, Hayakawa M, Matsumaru Y, Takigawa T, Koguchi Y, Miyamoto Y, Mizuno Y, Chida K, Akahane K, Tsuboi K, Sakae T, Sakurai H.
Organizer
16th International Conference on Solid State Dosimetry
Place of Presentation
Sydney, Australia
Year and Date
20100919-24