Study of the phoswich type scintillation fiber detector

Research Project

  • PDF
Project/Area Number 22860082
Research Category

Grant-in-Aid for Research Activity Start-up

Allocation TypeSingle-year Grants
Research Field Nuclear engineering
Research InstitutionNational Institute of Radiological Sciences

Principal Investigator

KOBA Yusuke  独立行政法人放射線医学総合研究所, 重粒子医科学センター, 研究員 (10583073)

Project Period (FY) 2010 – 2011
Keywordsフォスウィッチ型検出器 / 小型検出器
Research Abstract

We developed the small phoswich type detector for the measurements of the radiation quality in the heavy ion radiothrapy. The detection part of this detecor consists of the hollow type CaF2 crystal and liquid organic scintillator. And we lead out the scintillation using the optical fiber. This study has shown that this type detector is possible to identify charged particles and secondary gamma rays by the result of the experiment using accelerator. And it has shown that higher energy resolution of this detector is required for the measurement of the precise radiation quality.