2014 Fiscal Year Final Research Report
Development of an Intense Micro X-ray Source Using a Capillary Lens
Project/Area Number |
23300186
|
Research Category |
Grant-in-Aid for Scientific Research (B)
|
Allocation Type | Single-year Grants |
Section | 一般 |
Research Field |
Medical systems
|
Research Institution | Tokyo Institute of Technology |
Principal Investigator |
HASEGAWA Jun 東京工業大学, 総合理工学研究科(研究院), 准教授 (90302984)
|
Co-Investigator(Kenkyū-buntansha) |
OGURI Yoshiyuki 東京工業大学, 原子炉工学研究所, 教授 (90160829)
|
Project Period (FY) |
2011-04-01 – 2015-03-31
|
Keywords | 準単色X線 / マイクロX線蛍光分析 / キャピラリーレンズ |
Outline of Final Research Achievements |
We developed a novel micro X-ray fluorescence (micro-XRF) system based on proton-induced quasi-monochromatic X-ray. A micro X-ray beam was produced by focusing 8-keV X-ray, which was emitted from a copper thin foil bombarded by 2.5-MeV protons, by a poly-capillary lens. By using this X-ray beam, we performed a micro-XRF measurement and obtained a lower detection limit of 2.3 ng for cobalt deposited on a copper substrate. In addition, we succeeded in reconstructing the micro pattern of cobalt thin foil formed on the substrate with a spatial resolution of around 250 microns. An X-ray irradiation stand using 45-degree bending magnet was constructed for the trace element analysis for living biological samples. With this device, we successfully performed an in-vivo measurement of the cobalt concentration in a leaf of living duckweed floating on 500-ppm cobalt solution.
|
Free Research Field |
イオンビーム応用工学
|