2013 Fiscal Year Final Research Report
Imposed expiratory work of breathing and support sysytem
Project/Area Number |
23592286
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Multi-year Fund |
Section | 一般 |
Research Field |
Anesthesiology/Resuscitation studies
|
Research Institution | Osaka University |
Principal Investigator |
UCHIYAMA Akinori 大阪大学, 医学(系)研究科(研究院), 講師 (00324856)
|
Co-Investigator(Kenkyū-buntansha) |
FUJINO Yuji 大阪大学, 医学(系)研究科(研究院), 教授 (50252672)
IGUCHI Naoya 大阪大学, 医学部附属病院, 助教 (00372623)
|
Project Period (FY) |
2011 – 2013
|
Keywords | 人工呼吸 / 呼吸仕事量 / 呼気相 / 気管チューブ / 人工鼻 |
Research Abstract |
The resistance of the endotracheal tube (ETT), the heat and moisture exchanger (HME), and the ventilator may affect the patient's respiratory status. Although previous studies examined the inspiratory work of breathing (WOB), investigation of WOB in the expiratory phase is rare. Firstly, we estimated tracheal pressure at the tip of the ETT and calculated expiratory WOB imposed by the ETT, the HME, and the ventilator. As mean expiratory flow increased, imposed expiratory WOB increased. Levels of imposed expiratory WOB were affected by ETT diameter and ventilator mode. Secondly, in order to reduce imposed expiratory WOB, we developed a new version of expiratory support system, which was attached to the respiratory circuit of the ordinary ventilator. This system consisted of five electromagnetic valves and a piston syringe, which were regulated by the signals of airway pressure, respiratory flow, and tracheal pressure. It could reduce expiratory WOB due to a small size of ETT.
|
Research Products
(1 results)