2012 Fiscal Year Final Research Report
Assessment of Functionality of Medical Appliances Installed in Base-Isolated Hospitals and When Subjected to Vertical Motion
Project/Area Number |
23656341
|
Research Category |
Grant-in-Aid for Challenging Exploratory Research
|
Allocation Type | Multi-year Fund |
Research Field |
Building structures/materials
|
Research Institution | Kyoto University |
Principal Investigator |
|
Co-Investigator(Kenkyū-buntansha) |
MATSUMIYA Tomohiro 近畿大学, 建築学部, 講師 (20454639)
HOKI Kazuaki 北九州市立大学, 国際環境工学部, 講師 (70599026)
|
Project Period (FY) |
2011 – 2012
|
Keywords | 医療施設 / 免震 / 医療機器 / 鉛直動 / 機能維持 |
Research Abstract |
Examined in this research are the mechanism of amplification for the acceleration responseof medical appliances relative to the input acceleration and their effect onto the performance of the appliances. The appliances would sustain significant jumping and according damage when the maximum accelerations exceed 4g. Exceptions are those with flexible elements in the vertical direction, in which they would experience significant motion such as jumping even for the acceleration ranging 1 to 2g. A generic 3DOF model is developed to quantify the degree of acceleration amplification from the ground to the floor of the superstructure.
|