2014 Fiscal Year Final Research Report
Diagnostic reference levels for dental cone beam computed tomography
Project/Area Number |
24592846
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Multi-year Fund |
Section | 一般 |
Research Field |
Pathobiological dentistry/Dental radiology
|
Research Institution | The Nippon Dental University |
Principal Investigator |
SATO Kenji 日本歯科大学, 生命歯学部, 准教授 (50130670)
|
Project Period (FY) |
2012-04-01 – 2015-03-31
|
Keywords | 歯科用コーンビームCT / 診断参考レベル / 面積線量 / 空気カーマ |
Outline of Final Research Achievements |
To establish diagnostic reference levels for dental cone beam CT (dental CBCT), a statistical analysis was performed on the results from a questionnaire survey on dental CBCT at university dental hospitals in Japan. The third quartile dose-area products (DAP) and air kerma (DAP/FOV area) at rotation center were determined. For example, the third quartile at the protocols of 4×4 cm FOV, DAPs and air kermas for an adult, child, and all patiens are 470、420、460 mGy cm2, and 20、25、30 mGy, respectively.
|
Free Research Field |
放射線物理
|