2014 Fiscal Year Final Research Report
Development of calibration technique of image sensors toward realization of imaging power meter for high power lasers
Project/Area Number |
24760327
|
Research Category |
Grant-in-Aid for Young Scientists (B)
|
Allocation Type | Multi-year Fund |
Research Field |
Measurement engineering
|
Research Institution | National Institute of Advanced Industrial Science and Technology |
Principal Investigator |
NUMATA TAKAYUKI 独立行政法人産業技術総合研究所, 計測標準研究部門, 主任研究員 (60420288)
|
Project Period (FY) |
2012-04-01 – 2015-03-31
|
Keywords | レーザ / 感度 / プロファイラ / CCD / 近赤外 / 高出力 / 光減衰器 / 均一性 |
Outline of Final Research Achievements |
Responsivity evaluation technique of CCD image sensors and an optical attenuation method applicable for high power lasers were developed to realize an image sensor-based laser power meter. Investigation of pixel to pixel uniformity of responsivity of Si-CCD sensors revealed strong dependence on exposure time and laser wavelength. Proposed optical attenuator utilizing re-coupling of evanescent wave allowed attenuation of 10 W, φ3 mm laser beam down below -70 dB being easily controllable with the gap between two prisms. Incident beam profile was saved through the strong attenuation process. The attenuator and sensitivity-evaluated CCD were assembled to construct the imaging power meter and applied to laser power measurement.
|
Free Research Field |
レーザ放射計測
|