• Search Research Projects
  • Search Researchers
  • How to Use
  1. Back to project page

2014 Fiscal Year Final Research Report

Radiation dose analysis and development of various dose optimization methods in dual-energy CT

Research Project

  • PDF
Project/Area Number 24791279
Research Category

Grant-in-Aid for Young Scientists (B)

Allocation TypeMulti-year Fund
Research Field Radiation science
Research InstitutionKanazawa University

Principal Investigator

MATSUBARA Kosuke  金沢大学, 保健学系, 准教授 (30507372)

Project Period (FY) 2012-04-01 – 2015-03-31
KeywordsX線 / コンピュータ断層撮影 / 被ばく / 最適化 / エネルギー
Outline of Final Research Achievements

To optimize patient dose in dual-energy computed tomography (DECT), radiation dose and image quality, operating characteristics of automatic exposure control (AEC), and physical and clinical assessment of attenuation values in virtual non-contrast (VNC) images were evaluated. Although DECT provides images which have image quality inferior to single-energy CT (SECT), it can be performed with less radiation dose. The combination of AEC and DECT is unsuitable for obese patients because it tends to provide lower radiation dose and higher image noise levels than the combination of AEC and SECT. Attenuation in VNC images differs depending on the acquisition protocol, patient size, and contrast agent concentration.

Free Research Field

放射線防護学,X線CT技術学

URL: 

Published: 2016-06-03  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI

Powered by NII kakenhi