2014 Fiscal Year Final Research Report
Development of dose evaluation and treatment planning system of intensity modulated radiation therapy considering dose uncertainties
Project/Area Number |
24791297
|
Research Category |
Grant-in-Aid for Young Scientists (B)
|
Allocation Type | Multi-year Fund |
Research Field |
Radiation science
|
Research Institution | Niigata University |
Principal Investigator |
|
Project Period (FY) |
2012-04-01 – 2015-03-31
|
Keywords | 強度変調放射線治療 / 不確かさ / 放射線治療計画 / multi-leaf collimator |
Outline of Final Research Achievements |
Dose uncertainties in intensity modulated radiation therapy (IMRT) were analyzed qualitatively and quantitatively to create a numerical model to describe those dose uncertainties. We developed a method of creating a robust treatment plan against dose uncertainties due to patient setup error, and a method of quantifying and visualizing dose uncertainties due to error in multi-leaf collimator (MLC) modeling parameters appeared in commissioning of treatment planning system in dose distributions. Based on our developed method, a stable delivery of IMRT and maintaining its accuracy may be feasible even if there are some dose uncertainties which potentially affect to dose delivery accuracy of IMRT.
|
Free Research Field |
医学物理学
|