2015 Fiscal Year Final Research Report
Imaging of oxidant stress in brain tumor using an electromagnetic horn type ESR spectrometer including 3D MRI coil with cooling device
Project/Area Number |
25350534
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Multi-year Fund |
Section | 一般 |
Research Field |
Biomedical engineering/Biomaterial science and engineering
|
Research Institution | Oita University |
Principal Investigator |
Kamida Tohru 大分大学, 医学部, 准教授 (90315333)
|
Co-Investigator(Kenkyū-buntansha) |
KOBAYASHI Tadashi 大分大学, 工学部電気電子工学科, 名誉教授 (30100936)
Ohga Yasushi 大分大学, 工学部応用化学科, 教授 (60252508)
|
Project Period (FY) |
2013-04-01 – 2016-03-31
|
Keywords | 医用 / 生体画像 / ESR / 電磁ホーン / 脳腫瘍 / イメージング |
Outline of Final Research Achievements |
Imaging of a free radical in the living body is useful at diagnosis of the disease of oxidant stress origin, etc. In order to perform ESR measurement of a living specimen, X-band electromagnetic horn type ESR spectrometer including 3D MRI coil with cooling device capable of measuring both absorption spectrum and dispersion spectrum was developed, and compared with the conventional loop gap resonator in brain tumor model. Although ESR measurement from brain tumor was possible using a nitroxyl radical-labeled methionine for a tracer at the latter, non uniform magnetic field prevented ESR imaging in the former. X-band electromagnetic TM mode ESR spectrometer including 3D MRI coil with cooling device capable of measuring both absorption spectrum and dispersion spectrum was newly designed by application of the above-mentioned equipment.
|
Free Research Field |
総合領域
|