2015 Fiscal Year Final Research Report
The development of the gamma-ray spectrometer for electron measurement on fast ignition laser fusion experiment
Project/Area Number |
25630419
|
Research Category |
Grant-in-Aid for Challenging Exploratory Research
|
Allocation Type | Multi-year Fund |
Research Field |
Nuclear fusion studies
|
Research Institution | Osaka University |
Principal Investigator |
Arikawa Yasunobu 大阪大学, レーザーエネルギー学研究センタ, 講師 (90624255)
|
Project Period (FY) |
2013-04-01 – 2016-03-31
|
Keywords | X-ray spectrometer / fast ignition |
Outline of Final Research Achievements |
In this study we developed two kinds of high energy X-ray spectrometer. One is compton scattering X-ray spectrometer with the detective energy region of 0.5~5 MeV. The other is photo nuclear X-ray spectrometer with the detective energy region of 3~30 MeV. After this study we succeeded to develop two detectors as planed, and they were implemented to fast ignition laser fusion experiment. The electron energy spectrum was measured via bremsstrahlung X-ray spectrum by them, and we firstly found the generation of the electrons with the energy above 10 MeV which was origin of the low energy coupling efficiency of the fast ignition. The two x-ray spectrometer gave a great progress on the fast ignition laser fusion science.
|
Free Research Field |
fast ignition laser fusion
|