2016 Fiscal Year Final Research Report
Evaluation of the micromotion for immediately loaded implant
Project/Area Number |
25861893
|
Research Category |
Grant-in-Aid for Young Scientists (B)
|
Allocation Type | Multi-year Fund |
Research Field |
Dental engineering/Regenerative dentistry
|
Research Institution | Nara Medical University |
Principal Investigator |
HORITA Satoshi 奈良県立医科大学, 医学部, 助教 (60526713)
|
Project Period (FY) |
2013-04-01 – 2017-03-31
|
Keywords | 歯科インプラント / 即時荷重 / 微小動揺 / 有限要素法 |
Outline of Final Research Achievements |
The purpose of this study was to investigate the biomechanical behavior of immediately loaded implants in an edentulous mandible according to the All-on-Four concept.The results showed that mandibular fixed full-arch prostheses without cantilevers may result in a favorable reduction of the peri-implant bone strain during the healing period, compared with cantilevers.The maximum micromotion was within the acceptable limits for uneventful implant osseointegration according to the All-on-Four concept.Framework material did not play an important role in reducing the peri-implant bone strain and micromotion at the bone-implant interface.
|
Free Research Field |
歯科インプラント学
|