2016 Fiscal Year Final Research Report
Development of magnetically shielded roomless magnetocardiograph that enables early stage diagnosis of heart disease
Project/Area Number |
26350535
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Multi-year Fund |
Section | 一般 |
Research Field |
Medical systems
|
Research Institution | Iwate University |
Principal Investigator |
|
Research Collaborator |
IWAI Morio
MURAKAMI Tomohiko
KATO Daiki
|
Project Period (FY) |
2014-04-01 – 2017-03-31
|
Keywords | 心磁図 / SQUID磁束計 / 信号処理 / 信号源推定 / ノイズ除去 |
Outline of Final Research Achievements |
The purpose of this study is to develop a magnetically shielded roomless magnetocardiogram measurement system enabling early stage diagnosis of heart disease. In order to measure the magnetocardiogram with a high signal noise ratio outside the magnetically shielded room, magnetic noise was removed by hardware and software. The shielding factor by hardware was able to achieve about 27 dB in the ideal state. The shielding factor by software was able to achieve an average of about 33 dB. The magnetocardiogram waveform actually measured using this system outside the magnetically shielded room was able to realize a high signal noise ratio. Furthermore, it has become possible to create three-dimensional visualization software of the active part of the heart and to capture the activity of the heart three-dimensionally.
|
Free Research Field |
電子計測工学
|