2016 Fiscal Year Final Research Report
In-vivo monitoring using MEMS for the individual optimal design of implant prostheses
Project/Area Number |
26462927
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Multi-year Fund |
Section | 一般 |
Research Field |
Prosthodontics/ Dental materials science and
|
Research Institution | Kyushu University |
Principal Investigator |
|
Project Period (FY) |
2014-04-01 – 2017-03-31
|
Keywords | インプラント / MEMS |
Outline of Final Research Achievements |
The purpose of this study was to analyze stress in the time domain using MEMS sensor designed with a small circuit.As a result, 2.4 GHz Monopole Film Antenna for Implant Sensor that can be attached to the dental prosthesis was prepared, even in the presence of the buccal mucosa, if it is about 1.5 cm distance, it can transmit and receive outside the mouth. Next, the influence of the difference in attachment on the stress distribution on the implant and the displacement of the denture was evaluated . In the locator, the strain of the implant part was smaller than that of the other attachment and the displacement of the denture was the smallest.
|
Free Research Field |
インプラント
|