2017 Fiscal Year Final Research Report

Clarification of the threshold value of micromotion and the appropriate timing of starting of load to acquire osseointegration of immediate and early loaded implants

Research Project

  • PDF
Project/Area Number 26462973
Research Category

Grant-in-Aid for Scientific Research (C)

Allocation TypeMulti-year Fund
Section一般
Research Field Dental engineering/Regenerative dentistry
Research InstitutionNara Medical University

Principal Investigator

Kazuhiro Murakami  奈良県立医科大学, 医学部附属病院, 研究員 (60569078)

Co-Investigator(Kenkyū-buntansha) 山本 一彦  奈良県立医科大学, 医学部, 准教授 (20243842)
石田 純一  奈良県立医科大学, 医学部, 研究員 (40724638)
Project Period (FY) 2014-04-01 – 2018-03-31
Keywords歯科インプラント / 微小動揺 / 即時・早期荷重 / オッセオインテグレーション / 機能的動的荷重 / 骨形態計測 / レーザー変位計
Outline of Final Research Achievements

The micromotion (MM)of immediately loaded dental implants placed in the tibia of rabbits was monitored by a laser displacement sensor. The condition of the bone around the implants was histomorphometrically analyzed. The osteointegration of the implant could not be obtained in early loading even if a low loading value was applied, but the osseointegration could be obtained in delayed loading even if a high load value was applied. The initial loosening was not developed by loading at 5 days after implantation. In histomorphometrically analysis, bone contact rate showed the maximal value at 5 days after implantation and bone volume around implant showed the maximal value at immediate loading. From these results, the bone volume around implant can increase in the early loading, but the possibility of implant’s lost is increased. By setting a healing period for a while, the bone contacted rate increases, and the success rate also increases.

Free Research Field

生体力学

URL: 

Published: 2019-03-29  