2016 Fiscal Year Final Research Report
Development of fluorescent X-ray computed tomography using multi pinhole collimator
Project/Area Number |
26750142
|
Research Category |
Grant-in-Aid for Young Scientists (B)
|
Allocation Type | Multi-year Fund |
Research Field |
Biomedical engineering/Biomaterial science and engineering
|
Research Institution | Gunma University |
Principal Investigator |
|
Co-Investigator(Renkei-kenkyūsha) |
YUASA Tetsuya 山形大学, 大学院理工学研究科, 教授 (30240146)
SUNAGUCHI Naoki 北里大学, 医療衛生学部, 教授 (10197311)
|
Research Collaborator |
Sasaya Tenta
Murozuka Makoto
|
Project Period (FY) |
2014-04-01 – 2017-03-31
|
Keywords | 蛍光X線 / CT / 放射光X線 |
Outline of Final Research Achievements |
We propose a fluorescent x-ray computed tomography based on multiple pinholes (pFXCT) for preclinical molecular imaging. The proposed method simultaneously acquires multiple projections for a single view through the multi-pinhole collimator with a two-dimensional detector and a full-field volumetric beam to speed up data acquisition and to enhance the signal-to-noise ratio of projections. We constructed an pFXCT imaging system having fifteen pinholes at beamline ARNE-7A in KEK and performed preliminary experiments to investigate its imaging properties using physical phantoms including non-radioactive iodine imaging agent. The pFXCT attained a minimum detectable concentration of 38 I ug/ml required for in-vivo imaging at a spatial resolution of about 300 um for a data-acquisition time of 90 min less than an effective time of anesthesia, which suggests that preclinical molecular imaging is feasible with pFXCT.
|
Free Research Field |
医用画像工学
|