Development of a Membrane Oxygenator Combined with Pumping Function
Project/Area Number |
05671132
|
Research Category |
Grant-in-Aid for General Scientific Research (C)
|
Allocation Type | Single-year Grants |
Research Field |
Thoracic surgery
|
Research Institution | Kitasato University School of Medicine |
Principal Investigator |
TOMINAGA Seiichi Kitasato Univ., Sch. of Medicine, Associate Professor, 医学部, 助教授 (90050450)
|
Project Period (FY) |
1993 – 1994
|
Project Status |
Completed (Fiscal Year 1994)
|
Budget Amount *help |
¥2,100,000 (Direct Cost: ¥2,100,000)
Fiscal Year 1994: ¥1,000,000 (Direct Cost: ¥1,000,000)
Fiscal Year 1993: ¥1,100,000 (Direct Cost: ¥1,100,000)
|
Keywords | Membrane Oxygenator / Extracorporeal Circulation / Pumping Function / Silicone Tube / 体外循環 / 気体圧駆動膜型人工心肺 / ポンピング機能を有する人工肺 / 人工肺 |
Research Abstract |
A new membrane oxygenator combined with pumping function was designed to shorten the extracorporeal circulation circuit and to increase the efficiency of oxygenation. At a polycarbonate cylinder, 1500 silicone tubes were bundled and packed among elastic silicone filler. In both inflow and outflow sides of the cylinder, ball valves were equipped. The total available area of the membrane was 7.4*10^<-2> m^2. The membrane was pneumatically driven with a reciprocating pump in performance tests, in which 100% oxygen circulated in the gas side. The mock test shows that the partial pressure of oxygen increases from 0.7*10^4 Pa at the inlet to 7*10^4 Pa at the outlet of the oxygenator when 1.7*10^<-6> m^3 s^<-1> saline solution is pumped.
|