Sensorless measurement of biometric information for enhancing QOL in magnetically levitated ventricular assist device
Project/Area Number |
17H04909
|
Research Category |
Grant-in-Aid for Young Scientists (A)
|
Allocation Type | Single-year Grants |
Research Field |
Dynamics/Control
|
Research Institution | Tokyo Institute of Technology |
Principal Investigator |
|
Project Period (FY) |
2017-04-01 – 2020-03-31
|
Project Status |
Completed (Fiscal Year 2019)
|
Budget Amount *help |
¥23,920,000 (Direct Cost: ¥18,400,000、Indirect Cost: ¥5,520,000)
Fiscal Year 2019: ¥8,190,000 (Direct Cost: ¥6,300,000、Indirect Cost: ¥1,890,000)
Fiscal Year 2018: ¥6,760,000 (Direct Cost: ¥5,200,000、Indirect Cost: ¥1,560,000)
Fiscal Year 2017: ¥8,970,000 (Direct Cost: ¥6,900,000、Indirect Cost: ¥2,070,000)
|
Keywords | 磁気浮上型補助人工心臓 / 血栓検出 / 心拍同期制御 / 外乱オブザーバ / 補助人工心臓 / 磁気浮上 / センサレス計測 / 生体情報 / 流量制御 / 協調制御 / 人工心臓 / 磁気浮上システム / 血栓 / 血液粘度 |
Outline of Final Research Achievements |
In the field of ventricular assist devices, pump thrombus and control of appropriate flow rate have been issues to be solved. In this study, we have developed a method for detecting the pump thrombus from the phase difference between the current and displacement. Furthermore, we have also developed motor speed control system which can synchronize with the pulsatile of failure heart by using disturbance observer built in the magnetically levitated system.
|
Academic Significance and Societal Importance of the Research Achievements |
本研究は,人工心臓装着患者の生存率向上および生活の質(QOL)向上に繋がる.特に,追加のセンサを用いずに制御系の変更のみで実装可能な点で,既存の磁気浮上型補助人工心臓にも適用可能であり,汎用性が高い.更に,磁気浮上システムを用いた生体信号のさんサレス計測という,新しい応用分野の発展にもつながる点で,学術的意義が大きい.
|