• Search Research Projects
  • Search Researchers
  • How to Use
  1. Back to previous page

Power Doppler signal calibration between ultrasound machines by use of a capillary-flow phantom for pannus vascularity in rheumatoid finger joints

Research Project

Project/Area Number 24591437
Research Category

Grant-in-Aid for Scientific Research (C)

Allocation TypeMulti-year Fund
Section一般
Research Field 膠原病・アレルギー・感染症内科学
Research InstitutionHokkaido University

Principal Investigator

KAMISHIMA Tamotsu  北海道大学, 保健科学研究院, 教授 (10399868)

Co-Investigator(Kenkyū-buntansha) ATSUMI Tatsuya  北海道大学, 大学院医学研究科, 教授 (20301905)
TERAE Satoshi  北海道大学, 大学院医学研究科, 准教授 (40240634)
NISHIDA Mutsumi  北海道大学病院, 検査輸血部, 副技師長 (90404722)
Project Period (FY) 2012-04-01 – 2015-03-31
Project Status Completed (Fiscal Year 2014)
Budget Amount *help
¥2,860,000 (Direct Cost: ¥2,200,000、Indirect Cost: ¥660,000)
Fiscal Year 2014: ¥520,000 (Direct Cost: ¥400,000、Indirect Cost: ¥120,000)
Fiscal Year 2013: ¥1,300,000 (Direct Cost: ¥1,000,000、Indirect Cost: ¥300,000)
Fiscal Year 2012: ¥1,040,000 (Direct Cost: ¥800,000、Indirect Cost: ¥240,000)
Keywords関節超音波 / 関節リウマチ / 超音波 / 校正
Outline of Final Research Achievements

Ultrasound allows the detection and grading of inflammation in rheumatology. Despite these advantages of ultrasound in the management of rheumatoid patients, it is well known that there are significant machine-to-machine disagreements regarding signal quantification. In this study, we tried to calibrate the power Doppler (PD) signal of two models of ultrasound machines by using a capillary-flow phantom. In Aplio 500 and Avius, we found negative correlations between the PRF and the QPD index when the flow velocity was constant, and a positive correlation between flow velocity and the QPD index at constant PRF. The equation for the relationship of the PRF between Aplio 500 and Avius was: y = 0.023x + 0.36 [y = PRF of Avius (kHz), x = PRF of Aplio 500 (kHz)]. Our results suggested that the signal calibration of various models of ultrasound machines is possible by adjustment of the PRF setting.

Report

(4 results)
  • 2014 Annual Research Report   Final Research Report ( PDF )
  • 2013 Research-status Report
  • 2012 Research-status Report
  • Research Products

    (4 results)

All 2015 2014 Other

All Journal Article (1 results) (of which Peer Reviewed: 1 results,  Acknowledgement Compliant: 1 results) Presentation (2 results) Remarks (1 results)

  • [Journal Article] Power Doppler signal calibration between ultrasound machines using a capillary flow phantom for pannus vascularity in rheumatoid finger joints -a basic study2015

    • Author(s)
      Sakano R, Kamishima T, et al
    • Journal Title

      Radiological Physics and Technology

      Volume: 8 Issue: 1 Pages: 120-4

    • DOI

      10.1007/s12194-014-0299-5

    • Related Report
      2014 Annual Research Report
    • Peer Reviewed / Acknowledgement Compliant
  • [Presentation] Power Doppler signal calibration between ultrasound machines using a capillary flow phantom for pannus vascularity in rheumatoid finger joints -a basic study2014

    • Author(s)
      坂野稜典
    • Organizer
      第70回日本放射線技術学会総会学術大会
    • Place of Presentation
      パシフィコ横浜
    • Year and Date
      2014-04-01 – 2014-04-03
    • Related Report
      2014 Annual Research Report
  • [Presentation] Power Doppler signal calibration between ultrasound machines using a capillary flow phantom for pannus vascularity in rheumatoid finger joint -a basic study2014

    • Author(s)
      坂野 稜典
    • Organizer
      日本放射線技術学会 第70回総会学術大会
    • Place of Presentation
      パシフィコ横浜(横浜市)
    • Related Report
      2013 Research-status Report
  • [Remarks] 神島研究室ホームページ

    • URL

      http://www.hs.hokudai.ac.jp/kamishima/index.html

    • Related Report
      2013 Research-status Report

URL: 

Published: 2013-05-31   Modified: 2019-07-29  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI