キーワード | dose reconstruction / PET image / ion therapy |
研究実績の概要 |
A method for prediction of dose profiles from PET images for 11C and 15O ion beams were developed. The depth dose was measured for both ion beams with mono-energetic and poly-energetic conditions in the carbon therapy facility of the Heavy Ion Medical Accelerator in Chiba (HIMAC). The PET images were obtained for an irradiated polymethyl methacrylate (PMMA) phantom with both beams under the same conditions using a prototype PET system. The results of the mono-energetic beams were used for prediction of the dose profiles for the poly-energetic beams and the predicted dose profiles were in overall good agreement with the measured ones.
|
現在までの達成度 (区分) |
現在までの達成度 (区分)
3: やや遅れている
理由
For publication the accuracy and reproducibility of the data are essential. Therefore we are planning to perform more experiments. All experiments should be performed at the carbon therapy facility of the Heavy Ion Medical Accelerator in Chiba (HIMAC) using our prototype PET system and the number of experiments is so limited.
|
今後の研究の推進方策 |
The development and optimization of the method for prediction of dose profiles are going to be carried out for the measured data of an irradiated PMMA phantom with 12C ion beam and radioactive ion beams of 11C with SOBP.
|
次年度使用額が生じた理由 |
測定データの精度と再現性を評価するために追加実験をする必要があるが、この実験は量医研のHIMACのマシンタイムを利用して行う必要があり、多くの研究者が共同利用しているため、次の割り当ては来年度となっている。そして、この実験のために科研費予算を使用する必要があるため期間を延長する。
|