2002 Fiscal Year Final Research Report Summary
Development of real-time color intravascular ultrasound using Wavelet Analysis
Project/Area Number |
13670715
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Single-year Grants |
Section | 一般 |
Research Field |
Circulatory organs internal medicine
|
Research Institution | Yamaguchi University |
Principal Investigator |
FUJII Takashi Yamaguchi University School of Medicine, Associate Professor, 医学部, 助教授 (60228947)
|
Co-Investigator(Kenkyū-buntansha) |
HIRO Takafumi Yamaguchi University Hospital, Assistant Professor, 医学部附属病院, 助手 (10294638)
|
Project Period (FY) |
2001 – 2002
|
Keywords | Wavelet analysis / Intravascular Ultrasound / unstable plaque / Color mapping system / Coronary plaque |
Research Abstract |
Background and Objects: Wavelet analysis is a new mathematical model for assessing local changes in geometrical profile of time-series signals. We previously reported that Wavelet analysis of video-images of intravascular ultrasound (IVUS) might be useful for detecting lipid-rich area with atherosclerotic plaque in vivo. The purpose of this study is the development of real-time color intravascular ultrasound using Wavelet Analysis. Methods: IVUS signals were acquired from 32 formalin-fixed noncalcified atherosclerotic plaques from human necropsy by a 40 MHz IVUS catheter. Wavelet analysis of these images was performed with Daubechies-2 wavelet to obtain the correlation coefficient with the wavelet reconstructed over the x-y plane of the wavelet scale and the distance from IVUS catheter. Results: Histological examination revealed lipid-laden segment in 26 plaques. When performing a wavelet analysis, the bird eye view revealed an apparently different pattern in lipid-laden plaque compared to other types of plaque. Using this wavelet analysis, the lipid-laden plaque could be detected with sensitivity of 81%( 26/32) and specificity of 80%.
|
Research Products
(8 results)