2023 Fiscal Year Final Research Report
Quality assessment and quality control of digital breast tomosinthsis
Project/Area Number |
17K10375
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Multi-year Fund |
Section | 一般 |
Research Field |
Radiation science
|
Research Institution | Nippon Medical School |
Principal Investigator |
|
Project Period (FY) |
2017-04-01 – 2024-03-31
|
Keywords | 乳癌 / マンモグラフィ / トモシンセシス / 品質管理 / 画質評価 / 被ばく |
Outline of Final Research Achievements |
From a clinical perspective, we examined the performance, image optimization, and quality control of various breast tomosynthesis systems used clinically in Japan. We observed a reduction in radiation dose and improvement in CNR by changing the image reconstruction method from filtered back projection to iterative approximation, as well as the usefulness of iterative approximation in artifact evaluation. Because differences in DBT models and image reconstruction methods may cause changes in lesion visualization, it was suggested that it is necessary to understand the characteristics of image visualization ability.
|
Free Research Field |
放射線医学
|
Academic Significance and Societal Importance of the Research Achievements |
乳房トモシンセシス(DBT)は普及傾向にある一方でその品質管理が確立されていない現状があった。乳房トモシンセシスの複雑化したシステムの性能維持を保証する試験方法は簡便で定量的な画質評価が望まれている。DBTは従来のデジタルマンモグラフィと比較して被曝量が増加するために被ばく線量を最小にすることも重要であり、性能評価手順と品質管理手順の確立が急務である。DBTの品質管理方法が確立されていない現状で、臨床現場で実践可能な品質管理について検証を行う点は先駆的な取り組みである。
|