Outline of Final Research Achievements |
Proton beam therapy requires highly-precise image guidance in patient setup to ensure accurate dose delivery. Cone-beam CT (CBCT) is expected to play an important role to reduce uncertainties in patient setup. Particularly in particle therapy, the methodology to evaluate setup error considering dose distribution is to be established. This study aimed to develop the intelligent patient setup technique considering not only geometrical position of the patients but also dose distribution. Quantitative performance evaluation of CBCT image quality and mechanical accuracy of the system was performed. Accuracy of three-dimensional images registration between CBCT and reference CT images was evaluated. Moreover, the method to evaluate dose distribution under possible setup error was developed. Through these techniques feasibility of dose-guided radiotherapy has been established.
|
Free Research Field |
放射線治療
|