We developed a method for measuring ultra-trace level phosphate by online solid-phase extraction combined with HPLC. The detection limit (0.1μg-PO_4/L) was successfully improved with comparison to a conventional sensitive analytical method. In this method, ion par of phosphoantimonylmolybdenum blue and a quaternary ammonium ion was entrapped with a ODS mini-column, and the ion pair was eluted with CH_3CN/H_2O to quantified it by uv/vis detector. The advantages of this method are a wide detecting range (0.1-200μg-PO_4/L) and no disturbance with salinity.
|