Budget Amount *help |
¥27,430,000 (Direct Cost: ¥21,100,000、Indirect Cost: ¥6,330,000)
Fiscal Year 2014: ¥5,460,000 (Direct Cost: ¥4,200,000、Indirect Cost: ¥1,260,000)
Fiscal Year 2013: ¥14,040,000 (Direct Cost: ¥10,800,000、Indirect Cost: ¥3,240,000)
Fiscal Year 2012: ¥7,930,000 (Direct Cost: ¥6,100,000、Indirect Cost: ¥1,830,000)
|
Outline of Final Research Achievements |
In this study, we investigated on a multimodal ultra-low field MRI (ULF-MRI) system with optically pumped atomic magnetometers (OPAMs) as magnetic sensors to measure biomagnetic fields and MR signals. First, we carried out theoretical and experimental experiments on a K-Rb hybrid-type OPAM and demonstrate that the hybrid OPAM enabled to increase spatial homogeneity. In addition, we proposed a new method enabling simultaneous multi-location measurements by using the hybrid OPAM. Meanwhile, we proposed a new method for directly detecting magnetic resonance (MR) signals using a module-type OPAM in ULF-MRI. For the direct detection, we newly designed an actively shielded bias-field tuning coil to match the resonant frequencies between MR signals and OPAM. The results of numerical analyses indicate that actively shielded bias-field tuning coil can be designed to match the resonant frequencies of MR signals indicating its feasibility as an MR sensor toward a multimodal ULF-MRI.
|