Development of a graphite calorimeter for clinical electron beams
Project/Area Number |
24791365
|
Research Category |
Grant-in-Aid for Young Scientists (B)
|
Allocation Type | Multi-year Fund |
Research Field |
Radiation science
|
Research Institution | National Institute of Advanced Industrial Science and Technology |
Principal Investigator |
TANAKA Takahiro 独立行政法人産業技術総合研究所, 分析計測標準研究部門, 主任研究員 (30509667)
|
Project Period (FY) |
2012-04-01 – 2015-03-31
|
Project Status |
Completed (Fiscal Year 2014)
|
Budget Amount *help |
¥4,290,000 (Direct Cost: ¥3,300,000、Indirect Cost: ¥990,000)
Fiscal Year 2014: ¥130,000 (Direct Cost: ¥100,000、Indirect Cost: ¥30,000)
Fiscal Year 2013: ¥2,600,000 (Direct Cost: ¥2,000,000、Indirect Cost: ¥600,000)
Fiscal Year 2012: ¥1,560,000 (Direct Cost: ¥1,200,000、Indirect Cost: ¥360,000)
|
Keywords | グラファイトカロリーメータ / 治療用電子線 / 吸収線量 / 絶対計測 |
Outline of Final Research Achievements |
We have newly developed a graphite calorimeter for clinical electron beams. This graphite calorimeter has a thin graphite entrance window to detect clinical electron beams. The thin graphite window is specially designed for this graphite calorimeter, and is coated with a high density graphite layer by using CVD (chemical vapor deposition) method. Consequently, the core (absorber) of the calorimeter is successfully positioned shallower than the depth of dose maximum of 6 MeV electron beams. We tested the graphite calorimeter by external electrical power measurements. The results show that the graphite calorimeter can measure external power of 70 μW with a standard deviation of about 5 μW.
|