1990 Fiscal Year Final Research Report Summary

Development of a Precise Sample Position Control Device for Beam ExPeriments

Research Project

Project/Area Number 01850018
Research Category

Grant-in-Aid for Developmental Scientific Research (B).

Allocation TypeSingle-year Grants
Research Field 応用物理学一般(含航海学)
Research InstitutionTohoku University

Principal Investigator

KUSUNOKI Isao  Research Institure for Scientifec Measurements, Professor, 科学計測研究所, 教授 (30025390)

Co-Investigator(Kenkyū-buntansha) DOMEKI Hideo  ORIENTAL MOTOR Co., Ltd, Group Leader, 技術部, 主任
YUINE Tutomu  KOYO SEIKO Co., Ltd, Engineering Manager, 技術本部, 主担当員
YAMATO Yoshio  Research Institute for Scientific Measurements, Lecture, 科学計測研究所, 講師 (20006141)
Project Period (FY) 1989 – 1990
Keywordsmanipulator / ultra high vacuum / beam experiments / residual gas analysis
Research Abstract

A precisely controlable manipulator for use in molecular or ion beam scattering experiments was constructed by using uhv stepping motors, slide bearings and ball screws. The mechanical parts of the manipulator and the uhv aluminum vacuum container were constructed in the mechanical workshop of our institute, and the electronic parts were prepared by our laboratory. The movement test of the stages etc. By a personal computer and the vacuum test were also done in our laboratory.
The manipulator can control the sample position against the beam axis with six freedoms of ( X, Y, Z, rheta, phi, inclinatin ). The moving ranges of the sample were <plus-minus>22mm for the XY axis, <plus-minus>12mm for the Z axis, <plus-minus>7^゚ for the inclination, 360^゚ for both rheta and phi, and 830mm for vertical movement for sample exchange. The movement test for ( X, Y, Z, rheta) axis was finished without any problems. The other movements were now under testing. The vacuum test was carried out by a pumping system with a combination of a tarbo-molecular pump, an ion pump and a rotary pump. In the initial stage of pumping a large amount of gas was released, but the vacuum pressure reached into 10^<-9> Torr after several cycles of baking. In order to check the quality of the vacuum we analyzed the residual gas with a quadruple mass spectrometer. Some hydrocarbons and hydrogen chloride molecules in addition to the usual residual gas components (H_2, H_2O, CO, CO_2 ) were detected. The origin is unclear. Other problems are the outgas from the bearings during the motion and the increasing of temperature of the motor in operation. These problems should be solved in near future.

  • Research Products

    (8 results)

All Other

All Publications (8 results)

  • [Publications] 佐々木 俊明: "超高真空用6軸制御マニピュレ-タ-の試作" 東北大学科学計測研究所報告. 38. 139-145 (1989)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 楠 勲: "超高真空用ステッピングモ-タからの放出ガス" KEK Report(高エネルギ-研究所レポ-ト). 89ー2. 205-209 (1989)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 佐々木 俊明: "超高真空用6軸制御マニピュレ-タ-の試作" 技術研究会報告(分子科学研究所). 13. 36-41 (1990)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 楠 勲: "Stepping motor for use in UhV" Vacuum. 41. 1946-1947 (1991)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] T. Sakaki, M. Adachi, K. Aizawa, E. Ito, T. Saito, S. Yanagida, K. Minegishi and I. Kusunoki: "Construction of 6-Axis Controlled Manipulator for UHV application. ( in Japanese )" Bull. Res. Inst. Sci. Meas. Tohoku Univ.38. 139-145 (1989)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] I. Kusunoki, H. Satomi, M. Kobori and H. Domeki: "Outgas from the uhv stepping motors. ( in Japanese )" KEK Report. 89-2. 205-209 (1989)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] T. Sasaki, M. Adachi, K. Aizawa, E. Ito, T. Saito, S. Yanagida, K. Minegishi and i. Kusunoki: "Construction of 6-Axis Controlled Manipulator for UHV application. ( in Japanese )" Report of the meeting at IMS about technology. 13. 36-41 (1990)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] I. Kusunoki, H. Satomi, M. Kobori and H. Domeki: "Stepping motor for use in uhv." Vacuum. 41. 1946-1947 (1991)

    • Description
      「研究成果報告書概要(欧文)」より

URL: 