• Search Research Projects
  • Search Researchers
  • How to Use
  1. Back to project page

2000 Fiscal Year Final Research Report Summary

Development of a stereoscopic imaging system using fluorescent x-rays excited by synchrotron radiation

Research Project

Project/Area Number 10357009
Research Category

Grant-in-Aid for Scientific Research (A).

Allocation TypeSingle-year Grants
Section展開研究
Research Field Radiation science
Research InstitutionKYUSHU UNIVERSITY

Principal Investigator

TOYOFUKU Fukai  Kyushu University, School of Health Sciences, Professor, 医療技術短期大学部, 教授 (10117179)

Co-Investigator(Kenkyū-buntansha) YUASA Kenji  Kyushu University, Dental Hospital, Lecturer, 歯学部・附属病院, 講師 (40136510)
OHKI Masafumi  Kyushu University, School of Health Sciences, Professor, 医療技術短期大学部, 教授 (10160441)
TOKUMORI Kenji  Kyushu University, Fac. of Dentistry, Research assistant, 大学院・歯学研究院, 助手 (40253463)
KANDA Shigenobu  Kyushu University, Fac. of Dentistry, Professor, 大学院・歯学研究院, 教授 (20014003)
YOSHIURA Kazunori  Kyushu University, Fac.of Dentistry, Associate Professor, 大学院・歯学研究院, 助教授 (20210643)
Project Period (FY) 1998 – 2000
Keywordssynchrotron radiation / fluorescent x-ray / double-focus / stereo imaging
Research Abstract

The purpose of this study is to develop a double-focus fluorescent x-ray source excited by synchrotron radiation for stereoscopic imaging, aiming at real time visualization of three-dimensional structure of blood vessels..
(1) Development of a double-focus fluorescent x-ray source
The fluorescent x-rays were produced by exciting the metal targets by incident white x-ray from 6.5 GeV AR ring at Tsukuba. The intensities of fluorescent x-rays depend on several conditions such as incident x-ray spectrum, target metal, target angle, target thickness extraction angle and so forth. In order to obtain the optimum parameters, the computer calculations were performed. We developed the double-focus fluorescent x-ray source which is composed of two target boxes with rotating targets. They are placed with a distance of 5 to 30 cm apart along the beam line.
(2) Development of the stereoscopic imaging system
We developed data taking and video image processing system for stereoscopic imaging. Video signal … More s from the two image intensifier - TV camera systems are recorded on two digital video recorders. The video images are imported on a personal computer (Macintosh G4) and several pre-processing are performed frame by frame. Finally, stereoscopic video images are viewed using cross-eye method.
(3) Stereoscopic imaging experiments
Fluorescent x-rays, which range from about 30 keV to 75 keV can be generated by changing the target materials. In case of iodine contrast medium, fluorescent x-rays from Ba and Ce, having 32.2 and 34.7 keV K x-ray energies respectively, can be used for K-edge subtraction of iodine which has K absorption edge of 33.2 keV.In case of gadolinium, fluorescent x-rays from Ho and Yb, having 47.5 and 52.4 keV K x-ray energies respectively, can be used for gadolinium K-edge of 50.2 keV.Stereoscopic fluoroscopy has been performed by using phantoms which contain three different kind of contrast media (iodine, barium, and gadolinium).
Stereoscopic fluoroscopy has been performed by using phantoms which contain three different kind of contrast media (iodine, barium, and gadolinium). Less

  • Research Products

    (17 results)

All Other

All Publications (17 results)

  • [Publications] K.Tokumori,F.Toyofuku, et al.: "Monochromatic CT using fluorescent X-rays excited by synchrotron radiation"the 2000 World Congress of Medical Physics and Biomedical Engineering (CD-ROM). (2000)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 豊福不可依,徳森謙二: "放射光による単色X線CT"放射線技術学会雑誌. 56・6. 792-797 (2000)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 徳森謙二,豊福不可依 他: "蛍光X線源を用いた単色X線CT"医学物理. 20巻Sup.4. 258-259 (2000)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] F.Toyofuku,K.Tokumori, et al.: "Fan-beam monochromatic x-ray CT using fluorescent x-rays excited by synchrotron radiation"Proc.of SPIE. 3770. 48-56 (1999)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] K.Tokumori,F.Toyofuku, et al.: "Development of the photon-counting 256 channel CdTe line detector"Proc.of SPIE. 3770. 185-192 (1999)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] C.Uyama,K.Tokumori,F.Toyofuku et al.: "Proposal for medical application beamlines of the SPring-8"Proc.of SPIE. 3770. 193-204 (1999)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] F.Toyofuku, K.Tokumori: "Monochromatic x-ray CT using synchrotron radiation (in Japanese)"Japanese Journal of Radiological Technology. 56(6). 792-792 (2000)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] F.Toyofuku, K.Tokumori, S.Kanda, Y.Higashida, M.Ohki, T.Cho, K.Nishimura, K.Hyodo, M.Ando, and C.Uyama: "Fan-beam monochromatic x-ray CT using fluorescent x-rays excited by synchrotron radiation"Proc. of SPIE. Vol.3770. 48-56 (1999)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] K.Tokumori, F.Toyofuku, S.Kanda, S.Baba, Y.Mito, K.Hyodo, M.Ando, and C.Uyama: "Development of the photon-counting 256 channel CdTe line detector"Proc. of SPIE. Vol.3770. 185-192 (1999)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] C.Uyama, K.Tokumori, F.Toyofuku, H.Sugiyama, K.Hyodo, M.Ando, K.Nishimura, T.Takeda, K.Yamazaki, and M.Matsumoto: "Proposal for medical application beamlines of the SPring-8"Proc. of SPIE. Vol.3770. 193-204 (1999)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] T.Saito, H.Kudo, T.Takeda, Y.Itai, K.Tokumori, F.Toyofuku, K.Hyodo, M.Ando, K.Nishimura, and C.Uyama: "Three-dimensional monochromatic x-ray computed tomography using synchrotron radiation"Optical Engineering. 37(8). 2258-2268 (1998)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] F.Toyofuku, K.Tokumori, S.Kanda, K.Nishimura, K.Hyodo, M.Ando, C.Uyama: "Fluorescent x-ray source for diagnostic imaging studies"Medical Applications of Synchrotron Radiation, M.Ando and C.Uyama Eds., pp134-138 Springer-Verlag. (1998)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] K.Tokumori, F.Toyofuku, S.Kanda, K.Nishimura, K.Hyodo, M.Ando, and C.Uyama: "Accuracy of attenuation coefficients in monochromatic x-ray CT"Medical Applications of Synchrotron Radiation, M.Ando and C.Uyama Eds., pp175-178 Springer-Verlag. (1998)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] C.Uyama, T.Takeda, F.Toyofuku, K.Tokumori, M.Ando, K.Hyodo, Y.Itai, H.Shiwaku, K.Nishimura, and H.Kamitsubo: "Beamlines for medical applications on the Spring-8"Medical Applications of Synchrotron Radiation, M.Ando and C.Uyama Eds., pp149-158 Springer-Verlag. (1998)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] K.Nishimura, T.Fujisaki, K.Tokumori, F.Toyofuku, T.Takeda, K.Hyodo, M.Ando, and C.Uyama: "Scanning slit radiography system with synchrotron radiation"Medical Applications of Synchrotron Radiation, M.Ando and C.Uyama Eds., pp170-174 Springer-Verlag. (1998)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] F.Toyofuku, Y.Higashida, K.Tokumori, S.Kanda, C.Uyama, K.Hyodo, M.Ando: "Medical diagnostic technology using synchrotron radiation (in Japanese)"Hoshasen. 24(4). 51-60 (1998)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] K.Tokumori, F.Toyofuku, Y.Mito, S.Baba: "Application of CdTe array detector monochromatic x-ray CT (in Japanese)"Hoshasen. 24(2). 25-29 (1998)

    • Description
      「研究成果報告書概要(欧文)」より

URL: 

Published: 2002-03-26  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI

Powered by NII kakenhi