2016 Fiscal Year Final Research Report
Development of body insertion type SPECT/PET system
Project/Area Number |
15K15208
|
Research Category |
Grant-in-Aid for Challenging Exploratory Research
|
Allocation Type | Multi-year Fund |
Research Field |
Medical Physics and Radiological Technology
|
Research Institution | Tokyo City University |
Principal Investigator |
|
Project Period (FY) |
2015-04-01 – 2017-03-31
|
Keywords | SPECT / 悪性腫瘍診断 / 腹腔鏡手術 / とりこぼし |
Outline of Final Research Achievements |
Positron Emission Tomography (PET) and Single Photon Emission Computed Tomography (SPECT), which have been used as powerful tools to diagnose malignant tumors, have a limit to the intrinsic position resolution and detection capability. Moreover, laparoscopic operation is becoming mainstream in the field of tumor removal surgery. Therefore, new SPECT system that is applicable to the laparoscopic surgery having more than 5 mm position resolution is desirable. We have proposed a new body insertion type SPECT device constructed by 3 stick-type multi-pinhole gamma cameras. As a results of our studiy, it was shown that if there was imaging detector which resolution was less than 1mm, a tumor of 1 mm cubic was able to be detected by our proposed SPECT system.
|
Free Research Field |
放射線計測
|