• Search Research Projects
  • Search Researchers
  • How to Use
  1. Back to project page

2017 Fiscal Year Final Research Report

Development of a new ridge filter with honeycomb geometry for scanning system in particle radiotherapy

Research Project

  • PDF
Project/Area Number 15K19214
Research Category

Grant-in-Aid for Young Scientists (B)

Allocation TypeMulti-year Fund
Research Field Medical Physics and Radiological Technology
Research InstitutionNational Institutes for Quantum and Radiological Science and Technology

Principal Investigator

Tansho Ryohei  国立研究開発法人量子科学技術研究開発機構, 放射線医学総合研究所加速器工学部, 研究員 (80735126)

Project Period (FY) 2015-04-01 – 2018-03-31
Keywordsリッジフィルタ / スキャニング照射法 / ビーム微細化 / ハニカム型構造
Outline of Final Research Achievements

For particle radiotherapy, the current trend is to decrease lateral beam size to improve dose conformity in a tumor. We have developed a new ridge filter with 2-dimensional honeycomb geometry to simultaneously achieve both a decrease in lateral beam size and the desired energy modulation. We calculated dose distribution and found that the beam with smaller lateral size passing through a conventional ridge filter with 1-dimensional stepped structure cannot reproduce the designed dose distribution. However, we confirmed that using the new 2D ridge filter, the dose distribution was accurately reproduced. We actually manufactured the new ridge filter and also experimentally confirmed the reproducibility of dose distribution by the measurement.

Free Research Field

医学物理学

URL: 

Published: 2019-03-29  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI

Powered by NII kakenhi