• Search Research Projects
  • Search Researchers
  • How to Use
  1. Back to project page

2007 Fiscal Year Final Research Report Summary

Development of the dosimetry device with a magnetic levitation electrode ionization chamber which can measure a low level radioactive gas dose and neutron dose

Research Project

Project/Area Number 17560748
Research Category

Grant-in-Aid for Scientific Research (C)

Allocation TypeSingle-year Grants
Section一般
Research Field Nuclear engineering
Research InstitutionKyushu Sangyo University

Principal Investigator

KAWAGUCHI Toshiro  Kyushu Sangyo University, Department of Electrical Engineering, Faculty of Engineering, Professor (70069540)

Co-Investigator(Kenkyū-buntansha) IKEDA Nobuo  Kyusyu University, Graduate School of Engineering, Department of Applied Quantum Physics and Nuclear Engineering, Professor (70193208)
WAKABAYASHI Genichiro  Kyusyu University, Graduate School of Engineering, Department of Applied Quantum Physics and Nuclear Engineering, Assistant (90311852)
Project Period (FY) 2005 – 2007
Keywordsmagnetic levitated electrode ionization chamber / differential type / Tritium gas / ionization chamber / Neutron dosimetry / low level radioactive gas dose / space charge measurement
Research Abstract

A purpose of this study is to develop a high sensitivity magnetic levitation electrode ionization chamber which can measure a low level radioactive gas dose and a low level neutron dose. The study period was 2008 years from 2006.
At first, we developed a large-scale magnetically levitated electrode ionization chamber and examined a performance characteristic of the chamber with the volume of 8 liters. We confirmed that the ionization chamber can definitely measure low level radiation dose. We measured background dose under the condition that a device was put in an airtight container. We obtained the result that the opening of the chamber's shutter did not influence measurement of background. We measured also dose rate of background by the ionization chamber with the different volumes of 1, 8 and 14 liters simultaneously. Then, we obtained the same experimental results about magnitude and time fluctuation of the dose equivalent rate; even if ionization volumes were different. In measurement of 8 liters ionization chamber under time intervals 10, 30 and 60 minutes, a detection limit with reliability, 68% were 0.0054, 0.0027 and 0.0012 μSv/h, respectively.
At the next stage, we developed a differential type dosimetry device used a magnetic levitated electrode ionization chamber. This device is able to measure the space charge of 10-16A level. With this device, we measured very feeble tritium gas. Furthermore, by improving this device, we measured doses of a thermal neutron.
We announced these findings to Meeting of Atomic Energy Society of Japan. In addition, our articles was published by the RADIOISOTOPES(Vol.55, No.7) and the IEEE Transaction on Nuclear Science(Vol.53,No.4).

  • Research Products

    (9 results)

All 2008 2007 2006

All Journal Article (6 results) (of which Peer Reviewed: 2 results) Presentation (3 results)

  • [Journal Article] 磁気浮上電極電離箱を用いた空気イオン計測2007

    • Author(s)
      川口 俊郎, 二神 光次, 一木 博文
    • Journal Title

      九州産業大学工学部研究報告 43

      Pages: 50-56

    • Description
      「研究成果報告書概要(和文)」より
  • [Journal Article] Measuring ion in the atmosphere with magnetically levitated ionization Chamber2007

    • Author(s)
      T. Kawaguchi, F. Futagami, H. Ichiki
    • Journal Title

      Bulletin of the Faculty of Engineering Kyusyu Sangyo University Vol.43

      Pages: 49-56

    • Description
      「研究成果報告書概要(欧文)」より
  • [Journal Article] 大型磁気浮上電極電離箱の特性と環境放射線測定への応用2006

    • Author(s)
      川口 俊郎, 二神 光次, 的場 優, 他
    • Journal Title

      RADIOISOTOPES 55

      Pages: 361-369

    • Description
      「研究成果報告書概要(和文)」より
    • Peer Reviewed
  • [Journal Article] Low-Level Radiation Measurement System with Magnetically Levitated Electrode Ionization Chamber2006

    • Author(s)
      T.Kawaguchi, K.Futagami, M.Matoba
    • Journal Title

      IEEE Transaction on Nuclear Science 53

      Pages: 2276-2280

    • Description
      「研究成果報告書概要(和文)」より
    • Peer Reviewed
  • [Journal Article] Characteristics of a Large-sized Magnetically Levitated Electrode Ionization Chamber and an Application to Environmental Radiation Measurement2006

    • Author(s)
      Toshiro Kawaguchi, Koji Futagami, Masaru Matoba, Hirofumi Ichiki, Yoshitomo Utsunomiya
    • Journal Title

      RADIOISOTOPS Vol.55, No.7

      Pages: 361-369

    • Description
      「研究成果報告書概要(欧文)」より
  • [Journal Article] Low-Level Radiation Measurement System with Magnetically Levitated Electrode Ionization Chamber2006

    • Author(s)
      T. Kawaguchi, F. Futagami, M. Matoba, G. Wakabayashi, N. Ikeda, Y. Uozumi, N. Yamanaka, M. Kaneko
    • Journal Title

      IEEE Transaction on Nuclear Science Vol.53, No.4

      Pages: 2276-2280

    • Description
      「研究成果報告書概要(欧文)」より
  • [Presentation] Neutron dosimetry by Gd-loaded ionization chamber2008

    • Organizer
      2008 spring meeting of Atomic Energy Society of Japan
    • Year and Date
      2008-03-25
    • Description
      「研究成果報告書概要(欧文)」より
  • [Presentation] ガドリニウム装荷電離箱による中性子線量測定2008

    • Author(s)
      川口 俊郎, 惟永 和人, 一木 博文
    • Organizer
      日本原子力学会2008年春の大会
    • Place of Presentation
      大阪大学
    • Year and Date
      2008-03-24
    • Description
      「研究成果報告書概要(和文)」より
  • [Presentation] The background radiation measurement in a synchrotron light institution2007

    • Organizer
      2007 spring meeting of Atomic Energy Society of Japan
    • Year and Date
      2007-03-23
    • Description
      「研究成果報告書概要(欧文)」より

URL: 

Published: 2010-02-04  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI