2011 Fiscal Year Final Research Report
Study of the improvement in accuracy of a Dental fiberscope navigation system
Project/Area Number |
21592437
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Single-year Grants |
Section | 一般 |
Research Field |
Conservative dentistry
|
Research Institution | Tsurumi University |
Principal Investigator |
|
Project Period (FY) |
2009 – 2011
|
Keywords | 検査・診断システム / 内視鏡 / ナビゲーション |
Research Abstract |
We created a new dental fiberscope. However, due to the influence that focus depth and angle changes have on the scope, comprehending the device position and the precise target setting were difficult to ascertain consistently. Therefore, we developed a dental fiberscope that can be navigated. We verified the total accuracy of the dental fiberscope and resultant position accuracy of the navigation system. At the result, The navigation system error was 0.480mm±0.086mm(Difference X asis : 0.051mm±0.229mm, Y axis : 0.088mm±0.202mm, Z axis 0.254mm±0.260mm)
|
-
-
[Presentation] 根管治療時における歯科用ファイバースコープへのナビゲーションシステムの応用-第2報画像補正システムの導入-2011
Author(s)
山崎泰志, 小澤寿子, 重田優子, 井川知子, 木原琢也, 笠間慎太郎, 三島章, 細矢哲康, 小川匠, 小林馨
Organizer
第74回鶴見大学歯学
Place of Presentation
鶴見大学会館,横浜
Year and Date
2011-12-17
-
-
-