2012 Fiscal Year Final Research Report
Research for evaluation and improvement of proton delivery precision on new proton therapy system.
Project/Area Number |
22791227
|
Research Category |
Grant-in-Aid for Young Scientists (B)
|
Allocation Type | Single-year Grants |
Research Field |
Radiation science
|
Research Institution | Fujita Health University |
Principal Investigator |
NAOKI Hayashi 藤田保健衛生大学, 医療科学部, 講師 (00549884)
|
Project Period (FY) |
2010 – 2012
|
Keywords | 陽子線治療 / 品質管理 / 照射精度 / 線量測定 |
Research Abstract |
The purpose of this study was to establish the procedure for evaluating/improving the delivery precision using a new proton therapy system. Firstly, we investigated the radiologic behavior of proton beam by use of Monte Carlo simulation. Secondly, we evaluated several detectors for measuring the radiation dose from proton beam. As results, we estimated that the energy-loss of proton beam was happened at the irradiation-port of proton therapy system. Furthermore, we particularly made sure the characteristics of radiosensitive polymer gel dosimeter and radiochromic films. In conclusion, these tools are useful for proton beam dosimetry.
|