2017 Fiscal Year Final Research Report
Electromagnetic scanning forward viewing ultrasound endoscope equipped in a guide wire
Project/Area Number |
15H03028
|
Research Category |
Grant-in-Aid for Scientific Research (B)
|
Allocation Type | Single-year Grants |
Section | 一般 |
Research Field |
Medical systems
|
Research Institution | Tohoku University |
Principal Investigator |
Yoichi Haga 東北大学, 医工学研究科, 教授 (00282096)
|
Co-Investigator(Kenkyū-buntansha) |
西條 芳文 東北大学, 医工学研究科, 教授 (00292277)
松永 忠雄 東北大学, マイクロシステム融合研究開発センター, 助教 (00396540)
|
Project Period (FY) |
2015-04-01 – 2018-03-31
|
Keywords | 電磁駆動 / 超音波 / 内視鏡 / ガイドワイヤー / カテーテル / 微細加工技術 / 血管内治療 |
Outline of Final Research Achievements |
An electromagnetic scanning forward viewing ultrasound thin endoscope intended to be equipped in a guide wire has been developed for intravascular treatment. The endoscope consists of a resonator with an ultrasonic transducer and a driving coil. Driving of the electromagnetic resonator was confirmed using the driving coil, 50 um witdth silicon beam and a cylindrical shaped samarium cobalt magnet. Almost element technology has been established for fabrication of the device and equipment in the guide wire and ultrasound imaging will be realized by packaging in thin and small size.
|
Free Research Field |
医療機器開発
|